Operation
3.4.34 Hardware Menu
Advanced Menu Hardware Menu
This menu contains factory hardware installations. If the user reset to factory defaults, then they may
need to revisit this menu to enable their installed optional features.
Figure 76 Hardware Menu Screen
Model
Select the instrument model. Normally this only needs to be reset
when the configuration is corrupted. The selections available will
depend on licensing. It is not recommended to run an instrument
with firmware set to an incorrect model.
Front Panel Style
Choosing the incorrect front panel will result in the traffic lights
behaving inconsistently. Default is Aluminium.
Network Port
When Enabled indicates the instrument has a network port
installed. Default is Disabled.
Orifice Size
[Internal Pump Option Disabled]
Specify the input orifice for instruments without an internal pump.
Default is 1.08 for Standard range instruments and 1.50 for High
range instruments.
Internal Pump
When Enabled indicates the instrument has an Internal pump
installed. Default is Disabled.
Input 1 is CO2
When Enabled, analog input 1 is treated as a measurement of CO
2
.
This gas will appear on the Home screen and can be logged as the
second gas. Default is Disabled.
External Bkgnds
When Enabled the system will not perform automatic backgrounds
and it will not warn the user that it is not doing backgrounds. Rather
it is the responsibility of the operator to remotely trigger the
backgrounds. Default is Disabled.
IRSource Setpoint
Setpoint target for the IR Source voltage. At power up, the Source
Pot will slowly adjust until this setpoint is reached. This is normally
set by the factory.
Cooler Setpoint
Setpoint target for the Cooler Voltage in the IR Detector. At power
up, the Cooler Pot will slowly adjust until this setpoint is reached.
This is normally set by the factory