Serinus® 30 User Manual 3.3
Page 124
Note: Ensure that the instrument has been running for at least one hour before any calibration is
performed to ensure sufficient stability.
Note: Ensure units of measure are the same on both the barometer and instrument.
Equipment Required
Barometer
Procedure
1. Disconnect any external tubing connected to the rear of the instrument.
2. Open - Main Menu Calibration Menu Pressure Calibration - (read displayed note) - OK.
Note: Entering this menu will switch OFF the internal pump and place the valve sequencing on
hold; normal sampling will be interrupted.
3. Refer to Figure 115. Connect a barometer to the Sample Port.
Figure 115 Pressure Calibration (Internal Pump Option Only) Set-up
4. Edit - Vacuum Set Point - (read displayed instruction) - OK.
5. Wait 2 - 5 minutes and ensure the pressure reading on the barometer is stable.
6. Enter the barometer reading into the instrument - Accept.
7. Read displayed instructions - OK.
8. The pump should now stop automatically. Disconnect the barometer from the Sample Port.
9. Wait 2 - 5 minutes and enter the ambient barometer reading into the instrument - Accept.
Note: Both cell and ambient pressure sensors should now be displaying the current ambient
pressure and they should be the same value within 3 torr of each other.
10. Back - Pressure Calibration Menu - (read note) - OK.