Diagnostics Menu
Press/Temp/Flow Comp: ON
Diagnostic mode: Operate
Control Loop: Enabled
2.6 Transporting/Storage
Transporting the instrument should be done with great care. It is recommended that the packaging
the instrument was delivered in should be used when transporting or storing.
When transporting or storing the instrument the following points should be followed:
1. Turn OFF the instrument and allow it to cool down.
2. Remove all connections of pneumatic pipe, power cables and communication cables from the
instrument.
3. Refer to Figure 38. If storing over a long period (six months) turn the battery OFF by moving the
switch (S1) on the main controller PCA.
4. Remove the instrument from the rack.
5. Refer to Figure 39. Seal each pneumatic ports with a dust plug.
Figure 39 – Dust Plugs
6. Refer to Figure 37. If transporting the instrument, replace the travel screws.
CAUTION
Major damage to the instrument components may occur in transit if the
measurement cell is not secured in place with the supplied transit screws.
7. Refer to Figure 36. Remove the USB memory stick and pack with instrument.
8. Place the instrument back into a plastic bag with desiccant packs and seal the bag (ideally the bag
supplied upon delivery).