EasyManua.ls Logo

Ecotech Acoem Serinus 30 - Table 37 - Failure Bit Map (Positive Logic)

Ecotech Acoem Serinus 30
225 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix
Table 37 Failure Bit Map (Positive Logic)
Failure Bit
Meaning if set to 1
0
Flow sensor failure.
1
Instrument failure. Note that while the In
Maintenance mode reports as an instrument failure
with a red light on the front panel, for Bayern-
Hessen this particular error is merely a status
instead of a failure.
2
-
3
Lamp failure (S40 family only).
4
-
5
Cell heater failure (S30, S40 and S50 family only).
6
-
7
-
ST
Set the instrument mode.
Command Format
<STX>{ST}{< kkk>}<SP>{command}<ETC><bcc1><bcc2>
Where:
kkk Device’s Serial ID. This field is optional, but if provided it must be padded
with zeros to be 3 characters long. The value must match one of the
following: the instrument’s Bayern-Hessen ID, 000, or ??? (three
question marks).
Command M, N or K for Measure, Zero or Span mode.
bcc1 First byte of the block check calculation.
bcc2 Second byte of the block check calculation.
Device response
The device does not issue a response to this command.

Table of Contents

Related product manuals