Serinus® 30 User Manual 3.3
Page 184
Figure 157 Wiring Connections - CO
2
Probe to CO Analyser
Plumbing Connections
The CO
2
sensor does not restrict the flow in any way, so it can be connected either:
1. In-line between a sample manifold and the CO analyser; or
2. With its own sample pump and tube connections.
The direction of flow through the sensor is not critical. Use a piece of 1/4” O.D. Teflon tube to make
the connections. It is not usually necessary to pre-filter the sample entering the CO
2
sensor. The CO
2
sensor probe has its’ own replaceable filter. As the flow through the filter is diffuse and not high flow,
this filter should have a long service life in most installations. The filter is available as a spare part if
required.
It is important to check that the sensor probe is fully inserted into its black housing and ensure it is
sealed gas tight.
Figure 158 Example of installed CO
2
Sensor
Firmware Configuration
1. Press - (the green instrument status light), this will take the user to the home screen.
2. Press - (-99+) on the keypad. This will open the Advanced Menu.