Ecotech Acoem Serinus 30 - Equipment Required; Installation

Ecotech Acoem Serinus 30
225 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Optional Extras
Sensor output: 0 - 5 VDC
Sensor range: 0 - 2000 ppm (others available)
Power consumption: < 2.5 W
Operating temperature 0 - 40 °C
Operating pressure: 700 - 1300 hPa
Oper. humidity probe: 0 - 100% RH non condensing
Oper. humidity circuit board: 0 - 85% RH non condensing
Mounting: 2RU (88.9 mm high) 19” rack mounted panel suits
19” rack mounting
Inlet / outlet fittings: 1/4” quick-connect push-in fittings
8.8.3 Equipment Required
1/4” tubing such as Teflon tube will be required to connect the probe to a sample manifold.
8.8.4 Installation
Electrical Connections
The CO
2
sensor must be installed within an air conditioned room. The sensor output varies slightly with
temperature and pressure (refer specifications Section 8.8.2).
Connect the signal cable to the 25-way connector on the rear of the CO analyser as shown in
Figure 157.
Connect the power supply to a power outlet and switch the power ON.
Note: The plug pack supplied is universal AC input and 12 VDC output. The acceptable power
supply range is 11 - 20 VDC on +L terminal (refer to Figure 157). A second terminal labelled +H is
provided that allows the sensor to be powered from 24 VDC supply (18 - 30 VDC range) if required.
An analog (voltage) output is pre-wired with a cable and 25-pin male connector. The output is
0 - 5 VDC corresponding to a range of 0 - 2000 ppm. Other higher ranges are available from the
manufacturer consult Ecotech for information.