Serinus® 30 User Manual 3.3
The current temperature units of the instrument (Celsius, Fahrenheit
or Kelvin).
The temperature set point of the measurement cell. The factory
default is 50 °C.
Set Point (FLOW)
[Internal Pump Option]
The temperature set point of the flow block heater. The factory default
is 50 °C.
Displays current temperature of the measurement cell.
Displays current temperature of the CO Scrubber.
Flow Block
[Internal Pump Option]
Displays the current temperature of the flow block.
Displays the temperature of air inside the chassis, measured on the
main controller PCA.
Displays the temperature of the end-cap mirror plate on the
measurement cell.
Displays the humidity percentage of air inside the chassis, measured
on the main controller PCA board.
3.4.6 Pressure & Flow Menu
Main Menu → Analyser State Menu → Pressure & Flow Menu
Figure 48 – Pressure & Flow Menu Screen
Select the units that the pressure will be displayed in (torr, PSI, mBar,
ATM or kPa).
Current ambient pressure.
Current pressure within the measurement cell.
Select the units that the sample flow will be displayed in (slpm or L/min).
Flow Set Point
[Internal Pump Option]
Indicates the gas flow through the sample port of the instrument. The
value should be around 1.0 slpm. If there is an error with the sample flow,
it will read 0.00 slpm.