Serinus® 30 User Manual 3.3
Page 62
Temperature Units
The current temperature units of the instrument (Celsius, Fahrenheit
or Kelvin).
Set Point (CELL)
The temperature set point of the measurement cell. The factory
default is 50 °C.
Set Point (FLOW)
[Internal Pump Option]
The temperature set point of the flow block heater. The factory default
is 50 °C.
Cell
Displays current temperature of the measurement cell.
CO Scrubber
Displays current temperature of the CO Scrubber.
Flow Block
[Internal Pump Option]
Displays the current temperature of the flow block.
Chassis
Displays the temperature of air inside the chassis, measured on the
main controller PCA.
Mirror
Displays the temperature of the end-cap mirror plate on the
measurement cell.
Chassis Humidity
Displays the humidity percentage of air inside the chassis, measured
on the main controller PCA board.
3.4.6 Pressure & Flow Menu
Main Menu Analyser State Menu Pressure & Flow Menu
Figure 48 Pressure & Flow Menu Screen
Pressure Units
Select the units that the pressure will be displayed in (torr, PSI, mBar,
ATM or kPa).
Ambient
Current ambient pressure.
Cell
Current pressure within the measurement cell.
Flow Units
Select the units that the sample flow will be displayed in (slpm or L/min).
Flow Set Point
[Internal Pump Option]
The desired sample flow.
Sample Flow
Indicates the gas flow through the sample port of the instrument. The
value should be around 1.0 slpm. If there is an error with the sample flow,
it will read 0.00 slpm.