Serinus® 30 User Manual 3.3
Page 156
http://www.ecotech.com/downloads/firmware
Or by emailing Ecotech at service@ecotech.com or support@ecotech.com
To update the firmware from a USB memory stick, use the following procedure:
USB Memory Stick Update
1. Turn the instrument OFF.
2. Place the USB memory stick with the new firmware (ensure that firmware is placed in a folder
called FIRMWARE) in the front panel USB Port.
3. Enter the Bootloader (refer to Section 6.5).
4. Select option 3 (upgrade from USB memory stick), press 3 on the keypad.
5. Wait until the upgrade has completed.
6.5.4 Erase All Settings
This command is only required if the instruments firmware has become unstable due to corrupted
settings. To execute this command enter the Bootloader Menu (refer to Section 6.5) and press 4 on
the keypad.
6.5.5 Start Analyser
The start analyser command will simply initiate a firmware load by pressing 9 on the keypad from the
Bootloader Menu. It is generally used after a firmware upgrade.