3.4.28 Digital Output Menu
Main Menu → Communications Menu → Digital Output Menu
This allows the instrument to trigger external alarms in response to certain events. There are eight
different pins available, which will be set high during an associated event:
Figure 70 – Digital Output Menu Screen
Associates a state with a digital output. There are eight digital
outputs (the pin numbers are for the 25 pin connector). Each one
can have one of the associated states listed in Table 6. The pin
will be driven to the active state while the instrument state is
true.
Each pin can be set to be active High or Low. Active High means
that the pin will be pulled to 5 V when the associated event
occurs. Active Low means the pin will be pulled to 0 V when the
associated event occurs.
Table 6 – Digital Output States
No state (this state is never Active).
CO Scrubber heater fault.
Chassis temperature fault.
The USB memory stick is disconnected.