Serinus® 30 User Manual 3.3
Page 122
6. Refer to Figure 112. Connect a vacuum source to Exhaust Port of the instrument.
Figure 112 Full Pressure Calibration Set-up
7. Wait 2 - 5 minutes and ensure the pressure reading on the barometer has dropped and is stable.
8. Enter the barometer reading into the instrument - Accept.
9. Read displayed instructions - OK.
10. Turn OFF the vacuum source and wait 1 min or until the barometer reads ambient pressure.
11. Refer to Figure 113. Disconnect the vacuum source and barometer from Exhaust Port and Sample
Port.
Figure 113 Full Pressure Calibration Set-up (Ambient)
12. Wait 2 - 5 min, enter the ambient barometer reading into the instrument - Accept.
Note: Both cell and ambient pressure sensors should now be displaying the current ambient
pressure and they should be the same value within 3 torr of each other.
13. Back - Pressure Calibration Menu - (read note) - OK.