This command must be activated to safely remove the USB 
memory stick. 
Activating this will restart the instrument. 
 
3.4.14  Diagnostics Menu 
Main Menu → Service Menu → Diagnostics Menu 
 
Figure 56 – Diagnostics Menu Screen 
Internal Pump Menu 
[Internal Pump Option] 
On (default): Is used to compensate instrument measurements 
for environmental fluctuations that might affect readings 
(pressure, temperature and flow). 
Off: Is used only when running diagnostics. 
Enabled (default): Allows the instrument to automatically 
adjust digital pots and other outputs. 
Disabled: Prevents the instrument from changing most outputs 
so the service technician can manually control them. 
 
3.4.15  Digital Pots Menu 
Main Menu → Service Menu → Diagnostics Menu → Digital Pots Menu 
Digital pots are electronically controlled digital potentiometers used for adjustments to operations of 
the instrument. Each of the digital pots can go from 0 - 255. This menu should be accessed only during 
diagnostics. 
Unless the Control Loop is Disabled (refer to Section 3.4.14), changes to the pots may be modified by 
the instrument. This is intentional; some diagnostics are best done with instrument feedback and some 
are best done without.