Serinus® 30 User Manual 3.3
3.4.4 Status Menu
Main Menu → Analyser State Menu → Status Menu
The Status Menu presents a list of the current Pass/Fail status of the main components. During
warm-up, the status of some parameters will be a dashed line.
Figure 46 – Status Menu Screen
This field enters a screen with a log of all the events that the instrument
has performed. These events include errors and warnings. This log is
stored on the USB memory stick. The log is organised by month. When the
user enters this screen they will be prompted to enter the month for
which they wish to view events.
This field allows the user to display the list of current errors and warnings
on the screen.
This field is visible with the next service due date if the service is due
within the next two weeks.
Error if the system is “In Maintenance” (refer to Section 3.4.13).
Pass if the +5 V power supply is within the acceptable range.
Pass if the +12 V power supply is within the acceptable range.
Pass if the analog power supply is within the acceptable range (+12 V).
Pass if the analog power supply is within the acceptable range (-12 V).
Fail only if a problem is detected with the analog to digital conversion.
Pass if the cell heater temperature is within ± 10% of the heater set point.
Pass if the CO scrubber heater temperature is within ± 10% of 90 °C.
Pass if the mirror heater temperature is within ± 10% of set point.
Fail if the reference voltage is less than 0.5 V and the input pot is over 250,
which can be caused by the IR source failing.
Fail if the correlation wheel is not functioning.
Checks that the reference voltage is within acceptable limits 3.6 - 4.4 V.
Checks that the IR source voltage is 5 ± 0.5 V.
Pass if the system has an adequate electrical supply.