Serinus® 30 User Manual 3.3
Page 32
CO Detector PCA
The detector PCA amplifies the signal from the IR detector and it includes the trim pot to adjust the
detector gain.
Figure 14 CO Detector PCA
1.5.4.4 Pressure Sensor PCA
Refer to Figure 7 for the location of pressure sensor PCA. The pressure sensor PCA has an absolute-
pressure transducer that is mounted on the measurement cell and sealed via a gasket. This allows the
pressure sensor PCA to measure the sample air pressure in the measurement cell. This pressure
reading is used to verify sample flow and correct measurement readings for pressure variations.
Figure 15 Pressure Sensor PCA
1.5.4.5 Orifice and Sintered Filter
Refer to Figure 7 for the location of orifice and sintered filter. The critical orifice is made of a high
precision stainless steel and sapphire, which is located Inside the Kynar fitting on the end of the
measurement cell. The critical orifice is a simple device that operates at a set temperature and requires
minimal maintenance, it will passively keep the volume flow rate constant at a known value.
Figure 16 Orifice and Sintered Filter