43. Slowly remove the barometer from the Calibration Port,  allowing the  instrument to return to 
ambient pressure. 
44. Turn the manual stop valve 90° (Open). 
45. Open - Main Menu → Service Menu → Diagnostics Menu → Valve Menu. 
46. Enable - Valve Sequencing → Enabled. 
47. Open - Main Menu → Service Menu → Diagnostics Menu → Internal Pump Menu. 
48. Select - Pump Control → Start - Accept. 
6.4.5  CO-CO
2
 Converter Check  
The  internal  CO-CO
2
  converter  provides  a  source  of  continuous  CO  free  air  for  auto-zero  and 
background  functions.  Failure  of  the  converter  can  result  in  drift,  poor  sensitivity  near  zero  and 
continuous electronic zero adjusts.  
Note: Ensure that the instrument has been running for at least one hour before this procedure is 
performed. 
Equipment Required 
•  Span gas source 40ppm CO 
•  5/8” Spanner 
Procedure 
1.  Open - Main Menu → Calibration Menu. 
2.  Select - Cal. Type → Manual - Accept. 
3.  Select - Zero Source → Internal - Accept. 
4.  Select - Cal. Mode → Zero - Accept. 
5.  Allow the instrument to sample Zero air from the CO-CO
2
 for approximately 5 minutes. 
6.  Record the instrument response as the “Initial” value. 
 
Note: This value should be 0.00 ± 0.1 ppm. If it is not, a background should be initiated by initiating 
the background field within the calibration menu. The background will take approximately 5 
minutes to complete. 
 
7.  Connect  40  ppm  CO  span  source  to  the  Background  Air  Port  ensuring  the  inlet  pressure  is 
maintained at ambient (vented). 
8.  Allow the instrument to sample the span source for 5 minutes.  
9.  Record the instrument response as the “Challenge” value. 
10. Disconnect the span source. 
11. Open - Main Menu → Calibration Menu. 
12. Select - Cal. Type → Manual - Accept. 
13. Select - Zero Source → External - Accept. 
14. Select - Cal. Mode → Measure - Accept.