The mains power outlet must be protected by an earth leakage protection circuit.
Refer to Figure 28. Connect the instrument’s power cord into the instrument and mains
power outlet.
Figure 28 – Connect Power Cord
Refer to Figure 29. Turn ON the power switch.
Figure 29 – Turn ON Power Switch
2.3.2 Pneumatic Connections
The Serinus 30 has four pneumatic ports on the rear panel of the instrument, the Sample Port, the
Calibration Port, the Exhaust Port and the Background Air Port. All tubing and fittings used should
follow the instructions below:
Must be made of Teflon® FEP material, Kynar®, stainless steel, glass or any other suitably
inert material.
Sample line should be no more than two meters in length with 1/8” ID and 1/4” OD.
Sample inlet pressure should not exceed 5 kPa above ambient pressure.
Tubing must be cut squarely and any burrs removed.