Acclarix LX9 Series Diagnostic Ultrasound System User Manual
Transducers and Biopsy
- 110
-
4. Use distance measurement to obtain the depth of the target.
Note:
Center Line is not available on Intro-cavity transducers.
6.7 Needle Guide Bracket Cleaning and Sterilization
NOTE:
1. Use proper sterilization technique at all times when performing a biopsy.
2. Ensure that protective gloves are worn.
1. The needle guide bracket kits are not disinfected or sterilized before delivery. The operators must
clean and sterilize the needle guide bracket kits before the first use and after each use.
2. Inspect the bracket for damage such as cracks or breakage. If damage is evident, discontinue
use of bracket and contact your Edan representative for disposal guidance.
3. Sterilize the bracket before disposal or sending back to manufacturer for repair.
6.7.1. Cleaning
1. Wear sterile protective gloves to prevent infection.
2. Disconnect the needle guide bracket from the transducer after each use, and remove all visible
residues from the needle guide bracket using a small and soft-bristled brush or other similar
devices. Do the cleaning quickly before the needle guide bracket dries out.
3. Soak the needle guide bracket in the cleaning solution (Ethanol 75% or Isopropanol 70%) for at
least five minutes. Use a soft-bristled brush to clean the needle guide bracket during the soaking.
4. Take out the needle guide bracket from the cleanser and wipe all residues with a sterile cloth.
5. Let the bracket air dry, or dry the bracket with a sterile cloth.
6. If the bracket is not visually clean at the end of the cleaning steps, please repeat the cleaning
steps through step 3 to step 5.
7. Inspect the bracket to ensure that there is no damage. The bracket should be disposed of properly
when any damage is found.
6.7.2. Sterilization
1. Wear sterile protective gloves to prevent infection.
2. Disconnect the bracket from the transducer, and remove all visible residues from the bracket using
sterile cloth.
3. Clean and dry the bracket according to the methods in section 6.7.1 Cleaning.
4. Sterilize the bracket assembly by dynamic air removal steam sterilizer for at least four minutes at
132
o
C. Dry the bracket for at least 30 min after sterilization.
5. Inspect the bracket to ensure that there is no damage.
WARNING

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EDAN Acclarix LX9 Series and is the answer not in the manual?

EDAN Acclarix LX9 Series Specifications

General IconGeneral
BrandEDAN
ModelAcclarix LX9 Series
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals