- 231
-
Acclarix LX9 Series Diagnostic Ultrasound System User Manual
EMC Information
Appendix D EMC Information
Electromagnetic emissions
Guidance and manufacture’s declaration electromagnetic emission
The system is intended for use in the electromagnetic environment specified below; The customer or
the user of the system should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment guidance
RF emissions
CISPR 11
Group 1
The system uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF emissions
Class A
The system is suitable for use in all establishments other
than domestic and those directly connected to the public
low-voltage power supply network that supplies buildings
used for domestic purposes.
CISPR 11
Harmonic emissions
Class A
IEC 61000-3-2
Voltage
fluctuations/flicker
emissions
Complies
IEC61000-3-3
NOTE:
The EMISSIONS characteristics of the system make it suitable for use in industrial areas and
hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is
normally required) the system might not offer adequate protection to radio-frequency communication
services. The user might need to take mitigation measures, such as relocating or re-orienting the
equipment.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EDAN Acclarix LX9 Series and is the answer not in the manual?

EDAN Acclarix LX9 Series Specifications

General IconGeneral
BrandEDAN
ModelAcclarix LX9 Series
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals