Acclarix LX9 Series Diagnostic Ultrasound System User Manual
3.3 System Preparation
3.3.1. Battery Use
The system may come with two lithium-ion batteries depending on your order. The fully charged
batteries together can run the system for approximately 1 hour, depending on use. The batteries are
automatically charged when the system is plugged in.
1. If the system will remain unused for more than one week, charge the battery to at least 75%
capacity, take the battery out and store the system and battery separately.
2. During long term storage, the battery should be charged at least once every 6 months to ensure
battery capacity is more than 75%.
3. Only use Edan supplied battery.
4. Two batteries should be used together to power the system on.
To install the battery:
1. Find the battery compartment, located at the lower right side of the system(see figure 3-1) .
2. Unscrew the two screws on the battery door, and open the battery compartment.
3. Move the battery holder (see the figure below) up or down, and put two batteries inside
respectively. The side with labeling should face down when putting the battery in.