Acclarix LX9 Series Diagnostic Ultrasound System User Manual
5. Move the caliper to the end point.
Presses of the <Update> key will toggle between active calipers, allowing you to adjust the
position of start point and end point.
6. Press the <Set> key to fix the end point and complete the measurement.
2- Line Method:
1. Invoke the generic measurement function.
2. Select Distance-> Dist. 2 Line
3. The system pops up a line. Move the line to the target position, then adjust the angle of the
line and press the <Set>key to fix the first line.
4. Repeat step 3 to fix the second line, and system will display the distance between the two
lines.
Trace Method:
1. Invoke the generic measurement function.
2. Select Distance-> Length Trace.
3. Move the caliper to the start point.
4. Press the <Set> key to fix the start point.
5. Move the caliper to draw the trace between start point and end point.
Presses of the <Update> key will toggle between active calipers, allowing you to adjust the
position of start point and end point.
6. Press the <Set> key to fix the end point and complete the measurement.
8.1.1.2. Circumference/Area:
Circumference/Area measurement can measure the circumference and area of a closed region. There
are Ellipse, Trace and Spline methods.
Ellipse Method:
1. Invoke the generic measurement function.
2. Select Circ/Area-> Area Ellipse.
3. Move the caliper to the start point.
4. Press the <Set> key to fix the start point.
5. Move the caliper and press the <Set> key to fix the end point. An ellipse appears and can be
adjusted with the trackball.
Presses of the <Update> key will toggle between active calipers, allowing you to adjust the
ellipse diameter.
6. Press the <Set> key to complete the measurement.
Trace Method:
1. Invoke the generic measurement function.
2. Select Circ/Area-> Area Trace.
3. Move the caliper to the start point.
4. Press the <Set> key to fix the start point and start drawing the trace.
5. Moving the caliper to mark the trace outline.
6. Press the <Set> key to complete the trace.