Acclarix LX9 Series Diagnostic Ultrasound System User Manual
Measurements and Reports
Spline Method:
1. Invoke the generic measurement function.
2. Select Circ/Area-> Area Spline.
3. Move the caliper to the desired location.
4. Press the <Set> key to place the first point.
5. Repeat step 3~4 to place other points. A closed trace will be created after three points are
placed.
6. If a point is not placed satisfactorily, press Undo button to remove it. Each press of Undo
button removes one point.
7. Double press the <Set> key to complete the trace.
8.1.1.3. Angle
The Angle measurement can be done by 3-point method or 2-line method.
3- point Method:
1. Invoke the generic measurement function.
2. Select Angle-> Angle 3 Point.
3. Move the caliper to the vertex of the angle you are measuring, and press the <Set> key to fix
the first point.
4. Move the caliper to one end of the angle, and press the <Set> key to fix the second point.
5. Move the caliper to the other end of the angle.
Presses of the <Update> key will toggle between active calipers, allowing you to adjust the
position of the three points that make up the angle.
6. Press the <Set> key to fix the third point and complete the angle measurement.
2- line Method:
1. Invoke the generic measurement function.
2. Select Angle-> Angle 2 Line.
3. Place two lines following the 2-point distance method in section 8.1.1.1 Distance
measurement.
4. The angle between these two lines displays in the result window.
8.1.1.4. Volume
Volume can be measured with 3 Distances or 1 Ellipse+1 distance method.
3- Distance Method:
1. Invoke the generic measurement function.
2. Select Volume-> Volume 3 Dist.
3. Move the caliper to the start point.
4. Press the <Set> key to fix the start point.
5. Move the caliper to the end point.
Presses of the <Update> key will toggle between active calipers, allowing you to adjust the
position of start point and end point.
6. Press the <Set> key to fix the end point and complete the measurement of the first distance.
7. Repeat step 3~6 to complete the measurements of the second and third distance. Then the
Volume measurement result displays.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EDAN Acclarix LX9 Series and is the answer not in the manual?

EDAN Acclarix LX9 Series Specifications

General IconGeneral
BrandEDAN
ModelAcclarix LX9 Series
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals