- 234
-
Acclarix LX9 Series Diagnostic Ultrasound System User Manual
EMC Information
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the system is used exceeds
the applicable RF compliance level above, the system should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such
as reorienting or relocating the system.
b
Over the frequency range 150kHz to 80MHz, field strengths should be less than 3V/m.
c The ISM (industrial, scientific and medical) bands between 0.15 MHz and 80 MHz are 6.765
MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to
40.70 MHz. The amateur radio bands between 0.15 MHz and 80 MHz are 1.8 MHz to 2.0 MHz, 3.5
MHz to 4.0 MHz, 5.3 MHz to 5.4 MHz, 7 MHz to 7.3 MHz, 10.1 MHz to 10.15 MHz, 14 MHz to 14.2
MHz, 18.07 MHz to 18.17 MHz, 21.0 MHz to 21.4 MHz, 24.89 MHz to 24.99 MHz, 28.0 MHz to 29.7
MHz and 50.0 MHz to 54.0 MHz.
Table 1 Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications
equipment
Test
frequency
(MHz)
Band
a)
(MHz)
Service
a)
Modulation
b)
Maximum
power (W)
Distance
(m)
Immunity
test level
(V/m)
385
380-390
TETRA 400
Pulse
modulation
b)
18 Hz
1.8
0.3
27
450
430-470
GMRS 460,
FRS 460
FM
c)
± 5 kHz
deviation
1 kHz sine
2
0.3
28
710
704-787
LTE Band 13,
17
Pulse
modulation
b)
217 Hz
0.2
0.3
9
745
780
810
800-960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
b)
18 Hz
2
0.3
28
870
930
1720
1700-1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT; LTE
Band 1, 3, 4,
25; UMTS
Pulse
modulation
b)
217 Hz
2
0.3
28
1845
1970

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EDAN Acclarix LX9 Series and is the answer not in the manual?

EDAN Acclarix LX9 Series Specifications

General IconGeneral
BrandEDAN
ModelAcclarix LX9 Series
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals