EasyManua.ls Logo

EDAN F9 Express - User Manual

EDAN F9 Express
215 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
EDAN Agile PLM Electronic Signature Information
--Signatures related to this document and performed in EDAN Agile PLM.
文件名称(Document Name):F9说明书_英文
文件编号(Number):01.54.107779
版本(Version):2.9
产品型号(Product Model):F9;F9 Express
项目编码(Project Code):2040A000
签批信息(Signature):
作者(Originator) : 周 睿 (zhourui) 2018-10-15 15:36:46
审核人(Reviewers) : 刘 小倩 (liuxiaoqian) 2018-10-15 17:03:21
审核人(Reviewers) : 唐 林 (tanglin) 2018-10-16 09:10:49
审核人(Reviewers) : 明 镭 (minglei) 2018-10-16 13:54:54
审核人(Reviewers) : 张 冬雪 (zhangdongxue) 2018-10-19 12:04:28
审核人(Reviewers) : 占 皓 (zhanhao) 2018-10-29 09:12:01
审核人(Reviewers) : 余 俊煜 (yujunyu) 2018-10-16 18:54:43
批准人(Approvers) : 夏 欢欢 (xiahuanhuan) 2018-11-02 10:35:44
批准人(Approvers) : 吴 明花 (wuminghua) 2018-11-07 14:20:44
批准人(Approvers) : 冀 伟 (jiwei) 2018-11-02 15:35:48
版权©深圳市理邦精密仪器股份有限公司 (Copyright©Edan Instrument,Inc.)

Table of Contents

Questions and Answers

  • S
    Shelly HansonSep 6, 2025
    Why is the trace faint or missing on my EDAN F9 Express?
    • L
      Laura DeanSep 6, 2025
      If you're getting a faint trace or no trace at all on your EDAN Medical Equipment, make sure you're using the paper recommended by the manufacturer. Also, ensure that the paper drawer is locked by sliding it in until both latches are in position.
  • M
    Michelle EdwardsAug 22, 2025
    What to do if my EDAN F9 Express Medical Equipment has no display?
    • B
      Benjamin MorenoAug 22, 2025
      If your EDAN Medical Equipment shows no display, the issue might stem from a loose power cable. Ensure it's securely tightened. Alternatively, the fuse could be blown, requiring a replacement. Another possibility is that the battery is depleted; in this case, connect the equipment to an AC power supply.
  • M
    Megan RamirezSep 4, 2025
    Why is the trend a beeline on EDAN Medical Equipment?
    • B
      brownlindseySep 4, 2025
      If the trend on your EDAN Medical Equipment is a beeline, there may be a connector failure. Try moving or contacting the catheter. If the trend shows no fluctuation, change the intrauterine cable.
  • A
    Angela HopkinsSep 3, 2025
    Why only IUP peak but no baseline on EDAN Medical Equipment?
    • L
      Lisa ReynoldsSep 3, 2025
      If you're seeing only the IUP peak but no baseline on your EDAN Medical Equipment, the zero adjustment might be incorrect. Try zeroing the system.
  • E
    Eric NicholsSep 1, 2025
    How to fix a paper jam in EDAN F9 Express Medical Equipment?
    • C
      Christina Hernandez DDSSep 1, 2025
      If you're experiencing a paper jam with your EDAN Medical Equipment, it could be due to using the wrong type of paper or if the paper is dampened. Ensure you load the paper correctly and protect it from moisture.
  • A
    Andres GallagherAug 31, 2025
    What causes a doubtful FHR on EDAN F9 Express?
    • J
      Joseph HahnAug 31, 2025
      If you're getting a doubtful FHR reading on your EDAN Medical Equipment, it could be wrongly recording the maternal heart rate. Try changing the position of the ultrasound transducer. Another cause might be a poorly placed transducer recording mixed noise; adjust its position accordingly.
  • M
    Maria CooperAug 30, 2025
    Why is there ECG signal interference on my EDAN F9 Express Medical Equipment?
    • M
      Matthew MunozAug 30, 2025
      If you observe big ECG signal interference or a thick baseline on your EDAN Medical Equipment, it might be due to abnormal electrode placement or invalid electrodes. Check the electrodes' placement and validity. Also, ensure the cable connector is securely connected. Lastly, verify that the power socket has a standard ground wire.
  • T
    Thomas JimenezAug 28, 2025
    What to do if NIBP and SpO have no results on EDAN F9 Express?
    • M
      Matthew ArnoldAug 29, 2025
      If your EDAN Medical Equipment shows 'NIBP and SpO have no 2 results', check if the NIBP cuff is properly wrapped around the patient's arm. Extend the catheter and check its connection, if the NIBP cannot be inflated. Also, ensure the hose connector plug is securely connected to the NIBP socket and that the SpO transducer is well-connected to the SpO socket. If none of these resolve the issue, try shutting off the power and then switching it back on.
  • D
    Denise ChapmanAug 27, 2025
    Why is the trace inconstant on my EDAN F9 Express?
    • C
      curtisthomasAug 27, 2025
      If the EDAN Medical Equipment displays an inconstant trace or display, it could be due to several factors. If the patient is overweighted, monitor FHR with DECG. Ensure the ultrasound transducer is properly positioned for a better signal. Check for a loose belt and tighten it. Wipe off any superfluous aquasonic coupling gel, or use the recommended quantity if there is inadequate gel. Frequent fetal or maternal movements can also interfere; in these cases, delay monitoring or ask the patient to stay still.
  • N
    ngutierrezAug 27, 2025
    Why is the trace quality bad on my EDAN F9 Express?
    • M
      Mary WilliamsonAug 27, 2025
      If you're getting a bad trace quality or fluctuant TOCO baseline on your EDAN Medical Equipment, the belt might be too tight or too loose; try adjusting it. Also, if the belt has no elasticity, renew it. Maternal or frequent fetal movements can also cause this issue, so request the patient to calm down and stay still or delay the monitoring.

Summary

Chapter 1 Safety Guidance

Instruction for Safe Operation

Outlines critical safety instructions and guidelines for proper and safe use of the device.

Safety Precautions

Lists critical warnings and precautions to prevent injury or damage during instrument operation.

Chapter 2 Introducing the Monitor

Installation Guide

Provides step-by-step instructions for setting up and installing the monitor.

Chapter 4 Alarms

Chapter 6 Pre-Monitoring Preparation

Loading Recorder paper

Provides step-by-step instructions for loading the thermal printer paper correctly.

Switching On the Monitor

Details the procedure for powering on the monitor and checking its initial state.

Connecting Transducers

Provides instructions on how to properly connect various transducers to the monitor.

Chapter 7 Fetal Monitoring

Chapter 9 Maternal Monitoring (F9 Express)

Maternal NIBP Monitoring

Explains how to perform non-invasive blood pressure (NIBP) monitoring.

Chapter 12 Maintenance and Cleaning

Appendix 3 Troubleshooting

NIBP and SpO2 No Results

Troubleshooting steps when NIBP or SpO2 measurements yield no results.

EDAN F9 Express Specifications

General IconGeneral
Waterproof RatingIPX1
BatteryRechargeable lithium-ion battery
CategoryFetal and Maternal Monitor
FHR Measurement Range50 to 240 bpm
FHR Accuracy±1 bpm
Ultrasound Frequency1 MHz
Battery LifeApprox. 4 hours
PrintingThermal printer
Paper Speed1, 2, 3 cm/min
InterfacesUSB, RS-232
ParametersFHR, TOCO
Power Supply100-240 V~ 50/60 Hz
WeightApproximately 3.5 kg
PrintoutReal-time and recall printout
Display12.1-inch color TFT LCD screen, folding up to 90 degrees