Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block
Endress+Hauser 125
Adj. - Dens.
Adjustment
AUTO - OOS With this parameter a density adjustment can be carried out. The density
adjustment values will thus be recalculated and stored. This ensures that the
values dependent on density calculations (e.g. mass flow) are as accurate as
possible.
! Note!
Before carrying out a density adjustment, please refer to BA059D/06/en
"Promass 83 FOUNDATION Fieldbus Operating Instructions" where a
detailed description of the density adjustment is given.
Options:
CANCEL
START
Factory setting:
CANCEL
Adj. -
Dense.Set.Value 2
AUTO - OOS Use this parameter to enter the density setpoint value for the second fluid for
which you want to carry out field density adjustment.
User input:
5-digit floating-point number, incl. units
! Note!
• The preset density entered here should not vary from the actual fluid
density by a more than ±10%.
• The difference between the density setpoint values must be at least
0.2 kg/l.
• The associated unit is taken from the "System Units" parameter group.
Adj. - Meas.Fluid 2 AUTO - OOS In this parameter the current density of the second fluid is measured for
density adjustment.
Options:
CANCEL
START
Adj. -
Restore Original
AUTO - OOS With this parameter the original density coefficients determined at the
factory are restored.
Options:
NO
YES
Factory setting:
NO
Press.Corr. -
Pressure Mode
AUTO - OOS Use this parameter to configure an automatic pressure correction. In this
way, the effect of a pressure deviation between the calibration and process
pressures on the measured error for mass flow is compensated for, (see also
Operating Instructions BA065D, Accuracy Chapter).
Options:
OFF
FIX (A fixed process pressure for pressure correction is specified
→ "Press.Corr. - Pressure" parameter)
Factory setting:
OFF
"Flow" Transducer Block / base index 1400:
Parameter
Write access
with operating
mode
(MODE_BLK)
Description