3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus
138 Endress+Hauser
Main Line -
Assign
(continued)
Advanced options with optional software package
ADVANCED DIAGNOSTICS:
MASS FLOW DEVIATION
DENSITY DEVIATION
REFERENCE DENSITY DEVIATION
TEMPERATURE DEVIATION
TUBE DAMPING DEVIATION
ELECTRODYNAMIC SENSOR DEVIATION
OPERATING FREQUENCY FLUCTUATION DEVIATION
TUBE DAMPING FLUCTUATION DEVIATION
Main Line -
100%-Value
AUTO - OOS The entry is not active unless one of the following was selected in the
parameter "Main Line - Assign":
•MASS FLOW IN %
•VOLUME FLOW IN %
• CORRECTED VOLUME FLOW IN %
Use this parameter to define the flow value to be shown on the display as the
100% value.
User input:
5-digit floating-point number
Factory setting:
Depends on nominal diameter and country → Page 193 ff.
! Note!
The appropriate unit is taken from the corresponding parameter
("System Unit - Mass Flow", "System Unit - Volume Flow", or
"System Unit - Corrected Volume Flow"). ( → Page 114 ff.).
Main Line -
Format
AUTO - OOS Use this parameter to define the maximum number of places after the
decimal point displayed for the reading in the main line.
Options:
XXXXX. - XXXX.X - XXX.XX - XX.XXX -X.XXXX
Factory setting:
X.XXXX
! Note!
• Note that this setting only affects the reading as it appears on the display,
it has no influence on the accuracy of the system's calculations.
• The places after the decimal point as computed by the measuring device
cannot always be displayed, depending on this setting and the engineering
unit. In these instances an arrow appears on the display between the
measured value and the engineering unit (e.g. 1.2 → kg/h), indicating
that the measuring system is computing with more decimal places than
can be shown on the display.
"Display" Transducer Block/base index 1800
Parameter
Write access
with operating
mode
(MODE_BLK)
Description