3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus
150 Endress+Hauser
Info Line Mux -
Format
AUTO - OOS
! Note!
The option is not active unless a number was selected in the parameter "Info
Line Mux - Assign" ( → Page 148).
Use this parameter to define the maximum number of places after the
decimal point for the second value displayed in the information line.
Options:
XXXXX. - XXXX.X - XXX.XX - XX.XXX -X.XXXX
Factory setting:
X.XXXX
! Note!
• Note that this setting only affects the reading as it appears on the display,
it has no influence on the accuracy of the system's calculations.
• The places after the decimal point as computed by the measuring device
cannot always be displayed, depending on this setting and the engineering
unit. In these instances an arrow appears on the display between the
measured value and the engineering unit (e.g. 1.2 → kg/h), indicating
that the measuring system is computing with more decimal places than
can be shown on the display.
Info Line Mux -
Display Mode
AUTO - OOS
! Note!
The option is not active unless one of the following was selected in the
parameter "Info Line Mux - Assign" ( → Page 148):
• MASS FLOW BAR GRAPH IN %
• VOLUME FLOW BAR GRAPH IN %
• CORRECTED VOLUME FLOW BAR GRAPH IN %
Use this parameter to define the format of the bar graph.
Options:
STANDARD (Simple bar graph with 25 / 50 / 75% gradations and
integrated sign).
A0001258
SYMMETRY (Symmetrical bar graph for positive and negative directions of
flow, with –50 / 0 / +50% gradations and integrated sign).
A0001258
Factory setting:
STANDARD
Amp. Device Type read only This parameter is only used for service purposes.
"Display" Transducer Block/base index 1800
Parameter
Write access
with operating
mode
(MODE_BLK)
Description