3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus
156 Endress+Hauser
System Value -
Target Mass Flow
read only
! Note!
The option is not active unless one of the following was selected:
In the "Dens. - Function" parameter (Page 160)
% MASS / % VOLUME
FLEXIBLE and the % MASS 2D or % MASS 3D option in the
"Dens. - Mode" parameter (Page 163).
Use this parameter to display the currently measured mass flow of the target
fluid. Target fluid = carried material (e.g. lime powder).
Display:
5-digit floating-point number, including unit and sign
System Value -
Perc.Target Mass
Flow
read only
! Note!
The option is not active unless one of the following was selected:
In the "Dens. - Function" parameter (Page 160)
% MASS / % VOLUME
FLEXIBLE and the % MASS 2D or % MASS 3D option in the
"Dens. - Mode" parameter (Page 163).
In this parameter, the currently measured mass flow of the target fluid is
displayed as a % (of the overall mass flow). Target fluid = carried material
(e.g. lime powder).
Display:
5-digit floating-point number, including unit and sign
System Unit -
Mass Flow
AUTO - OOS Use this parameter to select the unit for displaying the mass.
Options:
Metric g; kg; t
US oz; lb; ton
Factory setting:
Depends on country (kg or US-lb)
! Note!
The unit of the totalizers is independent of your choice here. The unit for
each totalizer is selected separately for the totalizer in question.
System Value -
Target Volume Flow
read only
! Note!
The option is not active unless one of the following was selected:
In the "Dens. - Function" parameter (Page 160)
% MASS / % VOLUME
FLEXIBLE and the % MASS 2D or % MASS 3D option in the
"Dens. - Mode" parameter (Page 163).
Use this parameter to display the currently measured mass flow of the target
fluid. Target fluid = carried material (e.g. lime powder).
Display:
5-digit floating-point number, including unit and sign
"Calculated Density" Transducer Block/base index 2000