3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus
160 Endress+Hauser
Dens. - Function AUTO - OOS Use this parameter to select the desired density function which is used to
calculate special density values or the percentage proportion of components
in two-phase fluids.
User input:
OFF
% MASS / % VOLUME
%-BLACK LIQUOR
° BAUME > 1 SG
° BAUME < 1 SG
° API
° PLATO
° BALLING
° BRIX
FLEXIBLE
Factory setting:
OFF
Dens. - Ref.Density
Carrier Fluid
AUTO - OOS
! Note!
The option is not active unless one of the following was selected in the
parameter "Dens. - Function" ( → Page 160):
• % MASS / % VOLUME
• %-BLACK LIQUOR
Use this function to enter the reference density (density at reference temp.)
of the carrier fluid. This value is required for temperature-compensated
calculation of the target fluid content in a
two-phase fluid.
User input:
5-digit floating-point number, incl. units
Factory setting:
1.0000 kg/l
! Note!
• Carrier fluid = transporting liquid, (e.g. water)
Target fluid = material transported (e.g. lime powder)
• The appropriate unit is taken from the function UNIT REFERENCE
DENSITY (0421) ( → Page 117).
Dens. -
Lin.Exp.Coef.
Carrier
AUTO - OOS
! Note!
The option is not active unless one of the following was selected in the
parameter "Dens. - Function" ( → Page 160):
• % MASS / % VOLUME
• %-BLACK LIQUOR
Use this parameter to enter the fluid-specific linear expansion coefficient for
the carrier fluid for linear temperature curves. This value is required for
temperature-compensated calculation of the target fluid content in a
two-phase fluid.
User input:
5-digit floating-point number, including unit and sign
Factory setting:
0.5000 e
–3
[1/K]
"Calculated Density" Transducer Block/base index 2000
Parameter
Write access
with operating
mode
(MODE_BLK)
Description