Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block
Endress+Hauser 167
Visc. - Kin.Viscosity read only Use this function to view the calculated kinematic viscosity.
Display:
5-digit floating-point number, incl. unit.
(e.g. 5.5445 mm
2
/s; 1.4359 cSt)
Visc. -
Temp.Comp.Dyn.
Viscosity
read only Use this function to view the temperature-compensated dynamic viscosity.
Display:
5-digit floating-point number, incl. unit.
(e.g. 462.87 cP; 731.63 mPa·s etc.)
Visc. - Temp.Kin.
Viscosity
read only Use this function to view the temperature-compensated kinematic viscosity.
Display:
5-digit floating-point number, incl. unit.
(e.g. 5.5445 mm
2
/s; 1.4359 cSt)
Visc. - Unit Dyn.
Viscosity
AUTO - OOS
For selecting the unit required and displayed for measuring the dyn. viscosity.
Options:
•Pa·s
•mPa·s
•P
•cP
Factory setting:
cP
Visc. - Unit Kin.
Viscosity
AUTO - OOS
For selecting the unit required and displayed for measuring the kinem. viscosity.
Options:
•m
2
/s
•mm
2
/s
•St
•cSt
Factory setting:
cSt
Visc. -
Temp.Comp.Model
AUTO - OOS Use this function to select a formula model for temperature compensation.
Depending on the thermal behavior, you can select the model that most
accurately depicts the function → See Special Documentation on viscosity
measurement for Proline Promass 83I, SD102D.
Options:
Power law
Exponential
Polynomial
Factory setting:
Polynomial
Visc. -
Ref.Temperature
AUTO - OOS Use this function to enter the reference temperature which should act as the
reference point.
User input:
7-digit floating-point number
Factory setting:
0
Reference quantity:
°C
"Viscosity" Transducer Block/base index 2100
Parameter
Write access
with operating
mode
(MODE_BLK)
Description