Revision 1 168 9/10/2019
11.4 Instrument Specifications
Size
17”H x 19”W x 9”D (43 cm x 48 cm x 23 cm)
Weight
40 lb (18.2 kg)
Power Consumption
10 Watts
Universal 90-240VAC and 10-36VDC input.
Capable of being attached to redundant power
supply
Electrical
CSA C/US Certified
Classification
Class 1, Div 2, Group B,C&D, T3
Ambient Temp.
10-50°C (without enclosure)
Electronics
Dual processor design with intelligent
colorimetric sensor. To achieve greatest
sensitivity sensor uses 24 bit AD converter
equipped with temperature monitor on sensor
block. Sensor automatically calibrates on every
cycle to prevent drift.
Optical tape counter optimizes tape usage and
provides exact measurement of tape used.
User can program low tape alerts/alarms.
Real time clock for accurate event time
stamping.
Software
Windows based software for configuration and
monitoring.
Memory
Field programmable memory Firmware for unit
can be upgraded in the field.
Data Logging
Up to 10 months of phosgene concentration
data logs. 512 Kbytes non-volatile memory for
data logging and audit trail
Options
Sampling system, solar power, back panel, low
pressure sensor, auto-cal, and enclosure /
cabinets (temperature