4.8 The Configuration Menu
When CFNG (F4) is pressed on the Main page, the display appears as shown in Figure 4-11.
The function keys lead to four menus that are described in Sections 4.8.1 through 4.8.4
mV Run State Time
6 1S1 S1 115
Configure ?
F1 F2 F3 F4
GEN DIO AIO MSEQ
Figure 4-11: The Configuration Page
4.8.1 The Gen Menu
When GEN (F1) is selected on the Configuration page (Figure 4-12), the General menu is
presented (Figure 4-16).
mV Run State Time
6 1S1 S1 115
TAPE 224.95.05
ACK_AL OFF
L: mV OFF
L:ppm OFF
F1 F2 F3 F4
PGUP UP DOWN PGDN
Figure 4-12: The General Menu
TAPE - indicates the tape length that remains. The value can be changed (e.g. when the
tape is replaced) by pressing the EDIT key on the keypad. A query will be presented and the
appropriate tape length should be entered, followed by the ENTER key.
ACK_AL - This field is used to acknowledge all latched alarms. To acknowledge the latched
alarms, press the EDIT key and use the ENTER key to toggle the value to ON. Note that
any latched alarms will not unlatch if they are still in the alarm condition.
L mVThis field is used to set the mV on the display to corrected mV (indicated by mV on
the display) or live uncorrected live mV (indicated by LmV on the display). This function
(On/OFF) can be edited by pressing the EDIT key followed by the ENTER key.
L ppm - This field is used to set the concentration reading on the display to the last
calculated value (indicated by the absence of the L in front of the analysis units) or to the
current analysis live value (indicated by the presence of the L in front of the analysis units).
This function (On/OFF) can be edited by pressing the EDIT key followed by the ENTER key.
DAY - presents the day of the week and cannot be edited.
TIME - presents the present time and cannot be edited.
MODBUS ADD - Presents the current Modbus unit address. This parameter can be edited
by pressing the EDIT key, typing in a new value followed by the ENTER key.
CURRENT - Presents the sensor LED current in mA.
TEMP - indicates current temperature of the sensor block.
SENSOR - Presents the sensor firmware revision number and cannot be edited.
FIRM - indicates the analyzer firmware revision number and cannot be edited.