5.10.4.4 Differences between the Sample Tab and the
Calibration/Reference/Zero Gas Tabs
The Sample tab and the Calibration/Reference/Zero Gas Tabs are quite similar in format, but
the following differences between the two sets of tabs should be noted.
• Cal Gas Concentration indicates the concentration phosgene in the Calibration, Reference
and Zero gas.
• The Calibration and Zero Tabs do not include the Stream Calibration Used field.
The Alarms field contain only the High Alarm Enable and % Deviation Enable options.
• Runs to Average - indicates the number of consecutive runs to be averaged to obtain the
new gain factor during a calibration.
• Consecutive Runs - the Consecutive Runs field tells the analyzer how many times to run
the Gas Run before switching to the next run in the run queue. If no other run is enabled
and it is set up in the Automatic Sequence under the sequence tab, the analyzer will simply
continue running until another run is triggered either conditionally or scheduled. By default,
this field is set to 1.