Revision 1 52 9/10/2019
CAL OFF - This field is used to initiate a calibration run. To initiate the calibration run
press the EDIT button, and press ENTER to toggle the status from OFF to ON. The analyzer
will run the calibration according to the setup of the Calibration run (see Section 5.10.4) and
then revert back to the auto-sequence.
REF1 OFF - This field is used to initiate the Reference 1 run. To initiate the Reference
1 run press the EDIT button, and press ENTER to toggle the status from OFF to ON. The
analyzer will run the Reference 1 run according to the setup of the Reference 1 run (see
Section 5.10.4) and then revert back to the auto-sequence.
REF2 OFF - This field is used to initiate the Reference 2 run. To initiate the Reference
2 run press the EDIT button, and press ENTER to toggle the status from OFF to ON. The
analyzer will run the Reference 2 run according to the setup of the Reference 2 run (see
Section 5.10.4) and then revert back to the auto-sequence.