Revision B 120 Series Maternal/Fetal Monitor 5-7
2015590-001
Setup Procedures: User Setup Screens
IMPORTANT
EFFECTIVITY—All changes take effect immediately after a
selection is enacted in step six (6).
NOTE: While any setup screen is displayed, the primary labor parameters remain
displayed. While using a setup screen, if an alarm condition exists for a
maternal parameter that is not displayed, only an audio alarm (if enabled) is
issued. Upon exiting the setup screen, the visual alarm indication is shown.
Example. If the NBP Setup screen is displayed, the primary labor
parameters continue to be displayed as well as the maternal NBP area
of the screen. Under an alarm condition that affects NBP, both a visual
and audible (if enabled) alarm will be issued. Under an alarm
condition that affects MSpO
2 or MECG, only an audio alarm (if
enabled) is given while the NBP Setup screen remains displayed; once
the setup screen is exited, the visual alarm indication for MSpO
2 or
MECG will be shown.