7-16 120 Series Maternal/Fetal Monitor Revision B
2015590-001
Calibration: Display Check
Figure 7-3. Front Bezel
Testing
CAUTIONS
Following completion of the procedure:
LEAKAGE TEST—Perform a leakage and dielectric test on the
120 Series Monitor per applicable standards.
FUNCTIONAL TEST—Perform a functional test of the overall
monitor following the instructions provided in “Chapter 6,
Functional Checkout Procedure”.