6-6 120 Series Maternal/Fetal Monitor Revision B
2015590-001
Functional Checkout Procedure: Connecting the Simulator
Connecting the Simulator
This part of the procedure prepares the simulator for use.
NOTE: You must use a Model 325 Simulator for the functional checkout
procedure. (Monitors in the 120 Series do not work with Model 305
Simulators.)
1. Ensure the Model 325 Power switch is in the off position.
2. Connect the Model 325 Simulator’s power cord to the power receptacle on the
rear panel of the simulator; plug the other end of the power cord into a properly
grounded wall outlet of appropriate voltage.
3. Ensure the 120 Series Monitor
Power switch is in the off position.
4. Connect the simulator interconnect cable’s 50-pin end to the simulator’s
Fetal
Monitor
connector.
5. Insert the green plug of the FECG/MECG adapter cable, cat. no. (REF)
1442AAO, into the monitor’s
FECG/MECG connector.
6. Connect the sub-cables of the other end of the simulator interconnect cable into
the color-coded connectors on the monitor/adapter:
ECG, US, and UA.
7. Turn on the Model 325 Simulator. Verify that the green
Power indicator
illuminates.
8. Turn on the 120 Series Monitor.