Revision B 120 Series Maternal/Fetal Monitor 9-3
2015590-001
General Maintenance: Cleaning
Cleaning the Tocotransducer, Ultrasound Transducer, and MECG
Cables
CAUTIONS
ABRASION—Do not use abrasive cloth, sharp objects, or
abrasive cleaners.
ALCOHOL—Do not use Alcohol in cleaning solutions.
DISCONNECTION—Detach all transducers and cables from the
monitor.
IMMERSION—Only Nautilus Tocotransducers may be
immersed. Other transducers and cables may not be immersed or
held under running water.
1. Dampen a cloth or paper towel with one of the following products; then wring
out until only slightly wet:
îš‹ Sodium Hypochlorite 5.25 % (Bleach) diluted 10:1
îš‹ Cidex
îš‹ Sporicidin
îš‹ Soap and water
2. Rub soiled area until clean, taking care not to excessively wet the
tocotransducer diaphragm seal or the contact surface of the ultrasound
transducer.
3. Dry with a soft, dry cloth.