15-10 120 Series Maternal/Fetal Monitor Revision B
2015590-001
Troubleshooting: MECG Troubleshooting
MECG Troubleshooting
Table 15-7. MECG Troubleshooting
PROBLEM PROBABLE CAUSE POSSIBLE SOLUTION
MECG erratic or not functioning properly.
Transducer not properly connected to
monitor.
Electrodes not properly placed.
Clips not attached to electrodes properly.
Electrode gel dried.
Defective MECG cable.
Service required.
Selected lead providing
inadequate signal.
Ensure transducer is securely attached
to monitor.
Re-apply electrodes.
Check clip attachments.
Check electrodes and re-apply gel if
necessary.
Replace cable.
Call Biomedical Engineering
Department.
Change lead selection on MHR/P Setup
screen.
Dashes (– – –) shown in MHR/P area of
display.
Monitor unable to make a
determination due to insufficient signal.
Ensure patient is not asystolic.
Ensure electrodes are firmly secured to
patient.