5-28 120 Series Maternal/Fetal Monitor Revision B
2015590-001
Setup Procedures: Service Mode Screens
New Hospital
To define hospital-wide menu settings for the first time, or to re-define existing
settings:
1. Select NEW HOSPITAL in the Defaults field on the Install Options screen.
2. Exit the service mode.
3. Configure each of the user setup screens as desired.
4. Cycle power on the monitor.
Line Frequency
This field is used to define the line power frequency for the country in which the
monitor is being used. Use the Trim Knob control to alternate between 50 Hz and
60 Hz. (The monitor is factory set for the country in which it is being used.)
ECG Artifact Elimination
This field is used to enable/disable ECG artifact elimination which affects only the
direct FECG mode. Use the Trim Knob control to alternate between ON and OFF.
(The factory default setting is OFF.) When ON is selected, any new heart rate value
which differs by more than ±25 BPM from the previously calculated heart rate is not
printed on the strip chart paper. When OFF is selected, all heart rate values are
printed on the strip chart paper without regard to previous rates.
Scaling
This field is used to set the heart rate scale for the strip chart paper. Use the Trim
Knob control to alternate between 30–240 (BPM/cm) and 50–210 (BPM/cm). (The
monitor is factory set for the country in which it is being used.) The MSpO
2 trend
expanded scale is also affected by this setting. When the heart rate scale is 30–240,
the MSpO
2 expanded scale is 60–100 % (10%/cm). When the heart rate scale is 50–
210, the MSpO
2 expanded scale is 50–100 % (12.5%/cm).
Language
This field is used to set the language shown on the display and printed on the strip
chart recorder paper. Use the Trim Knob control to cycle through the available
options: ENGLISH, SPANISH, FRENCH, GERMAN, and ITALIAN. (The
monitor is factory set for the country in which is it being used.)
NBP One-Minute Interval
This field is used to enable/disable the one-minute interval selection on the maternal
BP Setup screen. Use the Trim Knob control to alternate between ON and OFF.
(The factory default setting is OFF.)
HR Offset
This field is used to enable/disable the secondary offset feature. Use the Trim Knob
control to cycle through the available options: 10 MIN, ON, and OFF. (The factory
default setting is 10 MIN.)