Revision B 120 Series Maternal/Fetal Monitor 5-19
2015590-001
Setup Procedures: FSpO2 Setup Screen
FSpO2 Setup Screen
Select the FSPO2 softkey to access the FSpO2 Setup screen (Figure 6-2).
IMPORTANT
SOFTKEY ACTIVATION—The Corometrics fetal patient
module cable must be plugged into the
FSpO2 connector in order to
activate the FSpO2 mode softkey.
Figure 5-9. FSpO
2 Setup Screen
Response Time
This field sets the response time or averaging mode (the time the monitor takes to
respond to changes in fetal oxygen saturation).
îš„ Slow: The FSpO2 display responds to step changes in the fetal saturation with
approximately 120 accepted pulses, or in about 50 seconds at 150 BPM.
îš„ Fast: The FSpO2 display responds to step changes in the fetal saturation with
approximately 30 accepted pulses, or in about 11 seconds at 150 BPM. (The
factory default setting is SLOW.)
When signal quality is poor, the response time becomes longer regardless of the
mode selected.
Print Interval
This setting determines the time interval for printing the FSpO2 values on the strip
chart paper.
165
172
30
47%
FSpO
2
EXIT
US
FECG
TOCO
RESPONSE TIME: SLOW
PRINT INTERVAL: 5 MIN
% O
2
TRACE: OFF
FSpO
2
SETUP