9-26 120 Series Maternal/Fetal Monitor Revision B
2015590-001
General Maintenance: Maternal Blood Pressure Tests
Dielectric (Hi-Pot) Tests
CAUTION
POWER OFF—Turn OFF the monitor prior to performing any of
the hi-pot tests.
Patient–to–AC-Line Using DC Voltage for One Minute
Connect the monitor to the hi-pot tester as shown in Figure 9-5.
NOTE: The hi-pot tester voltage is 5.656 kVdc.
Figure 9-5. Patient–to–AC-Line Using DC Voltage for One Minute (5.656 kVdc)