Revision B 120 Series Maternal/Fetal Monitor 9-27
2015590-001
General Maintenance: Maternal Blood Pressure Tests
 FECG: Connect an FECG test body to the monitor’s front panel ECG input. The
FECG test body can be the assembly leakage cable. Measure the voltage
breakdown.
F pass F fail
 MECG: Connect an ECG test body to the monitor’s front panel ECG input. The
MECG test body can be a shorted MECG cable or the assembly leakage cable.
Measure the voltage breakdown.
F pass F fail
 IUP: Connect an IUP test body to the monitor’s front panel UA input. The IUP
test body can be a SensorTip cable (wrapped in aluminum foil) or the assembly
leakage cable. Measure the voltage breakdown.
F pass F fail
 MSpO2: Connect an MSpO2 test body to the monitor’s front panel MSpO2 input.
The MSpO
2 test body can be a MSpO2 cable and sensor (wrapped in aluminum
foil) or the assembly leakage cable. Measure the voltage breakdown.
F pass F fail
 FSpO2: Connect an FSpO2 test body to the monitor’s front panel FSpO2 input.
The FSpO
2 test body should be an FSpO2 cable and sensor wrapped in
aluminum foil. Measure the voltage breakdown.
F pass F fail
 US: Connect an US test body to the monitor’s front panel US input. The US
test body should be an ultrasound transducer wrapped in aluminum foil.
Measure the voltage breakdown.
F pass F fail
 US2: Connect an US test body to the monitor’s front panel US2 input. The US
test body should be an ultrasound transducer wrapped in aluminum foil.
Measure the voltage breakdown.
F pass F fail