5-10 120 Series Maternal/Fetal Monitor Revision B
2015590-001
Setup Procedures: US/US2 Setup Screen
Volume
This field adjusts the volume for the FHR derived from the selected mode, US or
US2. This field works in conjunction with the front panel
Volume buttons.
FHR Audio Alarm
This field enables/disables the audio alarm function for FHR when derived from US/
US2.
îš‹ On: Visual and audible indications are provided during an FHR alarm
condition.
îš‹ Off: Only a visual indication is provided during an FHR alarm condition.
Alarm Volume
This field controls the alarm volume for Ultrasound alarms.