5-6 120 Series Maternal/Fetal Monitor Revision B
2015590-001
Setup Procedures: User Setup Screens
User Setup Screens
The 120 Series Monitor provides a variety of options that are all selected using setup
screens shown on the display. (The illustrations in this section are representative of
all possible features. Your monitor screens may vary.) All functions are performed
easily using the front panel Trim Knob control. The following setup screens are
selectable from the display:
îš‹ FECG
îš‹ US
îš‹ US2
îš‹ Maternal NBP
îš‹ MSpO2
îš‹ FSpO2
îš‹ MHR/P
îš‹ Master Alarm
îš‹ General
Refer to for Factory Default information.
Using the Trim Knob Control
General instructions for using the Trim Knob control follow:
1. To display a parameter setup screen, rotate the Trim Knob control until the bar
cursor highlights the title of the parameter (FECG, US, US2, NBP, MSpO
2,
FSpO
2, MECG, or PULSE). To access the Master Alarm Setup screen or the
General Setup screen, rotate the Trim Knob control until the bar cursor
highlights the
ALARMS softkey or the SETUP softkey, respectively, on the
bottom of the screen.
2. Press the Trim Knob control once to display the selected setup screen.
3. While the screen is displayed, rotate the Trim Knob control until the desired
field is highlighted.
4. Press the Trim Knob control once to activate the selected field. The cursor now
flashes.
5. Rotate the Trim Knob control in either direction to cycle through the available
choices for the field. Continuous rotation will wrap around through the choices.
6. When the desired selection is made for the field, press the Trim Knob control
once to enact the selection.
7. Repeat steps three (3) through six (6) until all desired settings have been made.
8. Rotate the Trim Knob control until the bar cursor highlights the
EXIT softkey on
the bottom of the screen. Press the Trim Knob control and the monitor returns
to normal operation.