6-34 120 Series Maternal/Fetal Monitor Revision B
2015590-001
Functional Checkout Procedure: Alarm Test
25. Wait ten seconds and verify the following:
îš‹ The alarm tone is once again emitted from the rear panel speaker.
îš‹ The MECG value flashes again.
26. Depress the monitor’s front panel
Alarm Silence pushbutton again.
27. Use the simulator’s
Manual Adjustment knob to decrease the MECG signal to
60 bpm. Verify the following on the monitor:
îš‹ The alarm tone is silenced.
îš‹ The MECG value no longer flashes.
îš‹ After 10 seconds, the two above conditions are still true.