6-38 120 Series Maternal/Fetal Monitor Revision B
2015590-001
Functional Checkout Procedure: Preventative Maintenance Inspection Report
page
9-18
Patient-to-Ground leakage for MSp02
(<50µA)
page
9-19
Patient-to-Line leakage for MSp02
(<50µA)
page
9-20
Patient-to-Ground leakage for FSp02
(<50µA)
page
9-21
Patient-to-Line leakage for FSp02
(<50µA)
page
9-22
Patient-to-Ground leakage for US
(
<50µA)
page
9-23
Patient-to-Line leakage for US
(
<50µA)
page
9-24
Patient-to-Ground leakage for US2
(
<50µA)
page
9-25
Patient-to-Line leakage for US2
(
<50µA)
page
9-25
Ground Continuity
Perform Functional Checkout in Chapters 6 & 8 of this Manual
Page
Description
Complete Max Actual Pass Fail N/A
page
8-3
Monitor Selt-Test
page
6-5
Front Panel Button Test
page
6-7
MECG Test
page
6-11
FECG Test
Leg Plate Test (Follow Manufacturer’s Directions)
page
6-16
Ultrasound Test
page
6-19
Fetal Movement Detection Test
page
6-16
Ultrasound Test
page
6-22
Uterine Activity Test
page
6-25
Tocotransducer Test
page
6-26
Strain Gauge Test (if used)
Page Description Complete Max Actual Pass Fail N/A