Revision B 120 Series Maternal/Fetal Monitor 7-13
2015590-001
Calibration: Display Check
10. Connect the front panel US2 harness cable to the Ultrasound Board at J6. Guide
the plastic cable bushing into the front-end enclosure until it snaps into place.
11. Connect the front panel US1 harness cable to the Ultrasound Board at J5. Snap
the bushing in place in the enclosure.
12. Connect the front panel UA harness cable to the Front End Motherboard at J7.
13. Connect the front panel FECG harness cable to the Front End Motherboard at
J6. Mount the cable’s ferrite bead in the retaining clip.
14. If BP is installed in the monitor, re-connect the blood pressure tubing to the
front panel connector board. Pay attention to the top/bottom labeling.
15. If MSpO
2 is installed in the monitor, connect the front panel MSpO2 harness
cable to the Oximeter Carrier Board at J6. Snap the bushing in place in the
enclosure. Mount the cable’s ferrite bead in the retaining clip.
16. If MECG is installed in the monitor, connect the front panel MECG harness
cable to the MECG Board at J7. Snap the bushing in place in the enclosure.
Mount the cable’s ferrite bead in the retaining clip.
17. If FSp0
2 is installed in the monitor, connect the front panel FSp02 harness cable
to the FSp0
2 board at J1. Snap the bushing in place in the enclosure. Mount the
cable’s ferrite bead in the retaining clip on the enclosure. Ensure that the second
ferrite bead is mounted on the back of the display shield.
18. Replace the hold-down cover for the front end boards and secure with all four
screws.
19. Re-connect the main switch power harness at the recorder divider.
20. Re-connect the recorder power supply cable harness to the Recorder Board at
J1.
21. Re-connect the recorder ribbon cable to the Recorder Board at J2.
22. Remove the new switch panel cable from its packaging. Connect one end to the
switch panel; connect the other end to the DSP Board at J4.
23. Connect the display interconnect cable to the DSP Board at J5.
24. If BP is installed in the monitor, ensure that the blood pressure tubing remains
routed through the notch at the top of the DSP Board.
25. Replace the monitor top cover and secure with all nine screws.
26. Replace the monitor power cord.
27. Re-configure the monitor using the Flasher Disk for a 3.5” x 4.5“ Display. The
Flasher disk re-configures the monitor software to work with the new larger
display. Follow the instructions included in the Flasher Display Configuration
Kit.