12-4 120 Series Maternal/Fetal Monitor Revision B
2015590-001
Upgrading a 120 Series Monitor: Communications Option Upgrade Kit
12. Connect applicable device(s) to the appropriate connectors on the monitor:
J101, J102, J103, and J104.
13. Although J104 is physically present on the Communications Board, it is only
supported in software as part of the Spectra Alerts option. The board is shipped
with a plug inserted in J104.
14. Re-connect the monitor power cord and turn on the monitor.
15. Test all functions of the devices connected to the monitor.
16. Fill out the Update Installation Verification Form included in the kit. Be sure to
include the monitor’s serial number and catalog number on the form. Follow
the instructions on the form to fax or mail the form to GE Medical Systems, Inc.