EasyManuals Logo

GE Corometrics 126 User Manual

GE Corometrics 126
440 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #403 background imageLoading...
Page #403 background image
Revision B 120 Series Maternal/Fetal Monitor A-5
2015590-001
Factory Defaults: Table of Defaults
NOTE: You must cycle power to reset factory defaults.
Install
Options
(Service)
Default Settings
Factory
Factory, Current, New
Hospital, Hospital-wide
Line Frequency United States: 60 Hz
International: 50 Hz
50 Hz, 60 Hz
ECG Artifact Elimination Off On, Off
Scaling United States: 30–240
BPM
International: 50–210
BPM
United States: 30–240 BPM
International: 50–210 BPM
Language Set according to shipping
destination
Set according to shipping
destination
NBP 1-Minute Interval On On, Off
HR (Heart Rate) Offset On with 10-min
Auto-Revert
Off, On, 10-min
Default TOCO Reference
10 relative units
5, 10, 15, 20, 25 relative
units
FM (Fetal Movement)
Remote Mark
On On, Off
COROLAN Address
Checking
Off On, Off
HBC (Heartbeat
Coincidence Enable)
On On, Off
VS (Vital Signs) Print
Interval
Real Time Real Time, Chart Style
Recorder Font Size Medium Small, Medium, Large
Fetal Alert/Alarm Off Off, Alarms, Alerts
Smart BP On On, Off
Table A-1. Summary of Factory Defaults
Setup
Screen
Field Description Factory Default Default Options Hospital Preference

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 126 and is the answer not in the manual?

GE Corometrics 126 Specifications

General IconGeneral
BrandGE
ModelCorometrics 126
CategoryMedical Equipment
LanguageEnglish

Related product manuals