Revision B 120 Series Maternal/Fetal Monitor B-3
2015590-001
Alarms Summary: Table of Alarms
NBP
An alarm setting (audio or high/low
limit) is turned off.
Alarm Defaults
Audio: on
Volume: 5
Systolic Limits: High = 160 mmHg,
Low = 90 BPM
Diastolic Limits: High = 90 mmHg,
Low = 50 mmHg
MAP Limits: High = 140 mmHg, Low
= 50 mmHg
MHR/P Limits: High = 120 BPM, Low
= 50 BPM
displays to the left of the
NBP mode title.
—
Systolic, diastolic, or MAP pressure
value (high or low) actively being
violated.
NBP numeric (systolic, diastolic,
or MAP) flashes.
Alternating high/low tones (if
audio enabled).
Malfunction with NBP circuitry, cuff,
or air hoses.
CHECK CUFF, LEAK, COMM,
MOTION
, WEAK SIGNAL, or
REPAIR message displays in NBP
area.
Alternating high/low tones (if
audio enabled).
MHR/P
a
An alarm setting (audio or high/low
limit) is turned off.
Alarm Defaults
Audio: on
Volume: 5
Limits: High = 120 BPM, Low = 50
BPM
Re-alarm: 120 sec
displays to the left of the
MHR/P mode title.
—
MHR/P limit (high or low) actively
being violated.
The tachycardia response time is < 8
seconds.
MHR/P numeric flashes. Alternating high/low tones (if
audio enabled).
Asystole. Flashing dashes “– – –” in place
of MHR/P numeric.
Alternating high/low tones (if
audio enabled).
MECG leads off. Flashing dashes “– – –” in place
of MHR/P numeric and
MECG
LEADS OFF
message displays
underneath.
Alternating high/low tones (if
audio enabled).
Table B-1. Summary of 120 Series Alarms
Type Condition Visible Advisory Audible Advisory