Chapter 9 Page no. 1118
JC-DR-A-150.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A150 - System Fuses
6 PROCEDURE
1. Refer to the appropriate illustration for the position of the fuse you want to change.
• For the Main Distribution Assembly fuse locations, refer to illustration 1.
• For the Generator Command Board fuse locations, refer to illustration 2.
• For the Generator Supply Command Board fuse locations, refer to illustration 3.
• For the Generator Main fuse locations, refer to illustration 4.
• For the Mains Distribution Board fuse locations, refer to illustration 5.
• For the PDU Board fuse locations, refer to illustration 6.
• For the MPCB Board fuse locations, refer to illustration 6.
2. If changing a fuse within the Generator, ensure that the components in the Generator are fully dis-
charged according to the steps described in section 5-1, Ensure Capacitors on Generator Boards are
Discharged on page 1110.
3. Change the fuse.
4. Power up the system and ensure that the system operates as expected without error.