Job Card D/R A150 - System Fuses
6 PROCEDURE
1. Refer to the appropriate illustration for the position of the fuse you want to change.
For the Main Distribution Assembly fuse locations, refer to illustration 1.
For the Generator Command Board fuse locations, refer to illustration 2.
For the Generator Supply Command Board fuse locations, refer to illustration 3.
For the Generator Main fuse locations, refer to illustration 4.
For the Mains Distribution Board fuse locations, refer to illustration 5.
For the PDU Board fuse locations, refer to illustration 6.
For the MPCB Board fuse locations, refer to illustration 6.
2. If changing a fuse within the Generator, ensure that the components in the Generator are fully dis-
charged according to the steps described in section 5-1, Ensure Capacitors on Generator Boards are
Discharged on page 1110.
3. Change the fuse.
4. Power up the system and ensure that the system operates as expected without error.